Products Description
The tube furnace features a compact design and incorporates a high-quality resistance wire heating element. It utilizes a double-shell structure and is equipped with a 30-program temperature control instrument. The system employs zero-trigger and SCR control technologies, ensuring precise temperature regulation. The furnace is constructed with polycrystalline alumina fiber material, and the tube ends are sealed with stainless steel flanges.
A gas nozzle, valve, and pressure gauge are installed on the stainless steel flange, allowing for controlled gas flow and pressure monitoring. The furnace can achieve a vacuum level of up to 10-3 Pa. It is designed with a balanced temperature field, resulting in low surface temperatures, rapid heating rates, and energy efficiency.
This tube furnace is an ideal product for high-temperature sintering, atmosphere control, reducing atmosphere experiments, CVD processes, and vacuum annealing. It is widely used in colleges and universities, research institutes, and industrial and mining enterprises for various high-temperature applications.
Technical Parameters
| Model | XW-DSP-1200℃-S-I | XW-DSP-1200℃-S-II |
| Power | 1.5 KW | 1.5 KW |
| Furnace tube size(mm) | Dia 25(OD) X700 /600mm | Dia50(OD)X700/600mm |
| Dimension | 340X290X415 | 340X290X415 |
| Power supply voltage | 220V | |
| Phase number | Single-phase | |
| Heating element | High-resistance wire | |
| Control mode | 1,30 sections of the program can only PID temperature regulation. | |
| 2,over-temperature protection, over-temperature protection | ||
| burnout or burnout furnace heating circuit automatically cut off , | ||
| (when the furnace temperature exceeds 1200 degrees or thermocouple burns out, | ||
| the AC relay on the main circuit will automatically disconnect the main circuit off, | ||
| the panel lights off oN, OFF lights, limited protection EAF). | ||
| 3, with 485 communication interface (when an optional software standard) | ||
| 4, with power protection function, namely, electrical power and then to start the stove, | ||
| the program does not start from a starting temperature rise, | ||
| but began to rise from the furnace power outage. | ||
| 5, the instrument with a temperature self-tuning features | ||
| Temperature control mode | Intelligent PID | |
| Control accuracy | +/- 1 ℃ | |
| Trigger | Zero trigger | |
| SCR | 106 / 16E Germany Simon Kang | |
| Max temperature | 1200 ℃ | |
| Rated temperature | 1100 ℃ | |
| Heating rate | ≤30 ℃/ Min (Available upon request modification) | |
| Heating length | 205mm | |
| Recommended heating rate | ≤15 ℃/ Min | |
| Vacuum degree | 1.General vacuum and drive gas out by:TW-1.5A | |
| (Optional) | 2.2XZ-2 Vacuum degree pumps can be able to get around 10Pa | |
| 3.XW-DSP-VS 0.1vacuum system can pumped 0.1Pa | ||
| (and equipped with a resistive vacuum gauge display) | ||
| Certification | CE | |
| Surrounding surface temperature | ≤45℃ | |