Products Description
1.XW-5V50mA Battery Analyzer
The XW-5V50mA is an 8-channel battery analyzer designed for testing small coin/button cells, with a current range of 0-50 mA and voltage range of 0-5 V. Each channel features an independent constant current (CC) and constant voltage (CV) source, programmable and controllable via dedicated computer software.
2.Versatile Testing Applications
This coin cell tester system supports a wide range of applications in battery testing, including electrode material research, battery performance evaluation, small-scale cell formation, capacity grading, and battery pack testing.
3.Compatibility
The analyzer is compatible with various coin/button cell types, such as lithium-ion batteries, Ni-MH batteries, zinc-manganese watch batteries, and other similar miniature electrochemical cells.
Technical Parameters
Item | TechnologyParameters | |||||
Power | AC 220V±10% / 50Hz | |||||
Input power | 25W | |||||
Resolution ratio | AD:16bit DA:16bit | |||||
Input impedence | ≥1 GΩ | |||||
Voltage | Constant voltage range control | 25mV~5V | ||||
Minimum discharge voltage | -5V | |||||
Precision | ± 0.05% of FS | |||||
Stability | ± 0.05% of FS | |||||
Current | Output range per channel | Range1:5uA~1mA,Range2: 1mA~25mA, Range 3: 25mA~50mA | ||||
Precision | ± 0.05% of FS | |||||
Constant voltage cut-off current | Range 1: 2uA,Range2: 50uA, Range 3: 100uA | |||||
Stability | ± 0.05% of FS | |||||
Power | Single channel maximum output power | 0.25W | ||||
Stability | ± 0.1% of FS | |||||
Time | Current response time | <500us (10%FS~90%FS) | ||||
Step time range | ≤(365*24)Hour/step | |||||
Time form support 00 :00 :00 :00 : 00 (h min s ms) | ||||||
Parameter record | Data recording condition | Minimum interval of time: 100ms | ||||
Minimum voltage interval:10mV | ||||||
Minimum current interval | ||||||
Range1: 2uA;Range2: 50uA, Range: 100uA | ||||||
Recording frequency | 10Hz | |||||
Charging | Charging mode | Constant current charging, constant voltage charging, constant current and constant voltage charging, constant power charging | ||||
Cut-off condition | Voltage, current, relative time, capacity -△V | |||||
Discharging | Discharging mode | Constant current discharge, constantvoltage discharge, constant current and constant voltagedischarge,constant power discharge, constant resistance discharge | ||||
Cut-off condition | Voltage, current, relative time, capacity -△V | |||||
Burst Mode | Charging | Constant current mode, constant power mode | ||||
Discharging | Constant current mode, constant power mode | |||||
Minimum pulse width | 500ms | |||||
Pulse number | A single pulse step supports 32 different pulses | |||||
Continuous switching of charge and discharge | A pulsed step can achieve continuous switching from charge to discharge | |||||
Cut-off condition | Voltage, relative time | |||||
DCIR test | Support custom fetch point for DCIR calculation | |||||
Circle | Cycle test range | 1~65535 | ||||
Single cycle step number | 254 | |||||
Nested loop | With nested loop function, maximum support 3 layer nesting | |||||
Protect | ● Disconnection with voltage data protection | |||||
●Offline testing function | ||||||
● It can set safety protection condition, set parameters include: voltage limit, voltage limit, current limit, current limit, capacity limit, delay time, with anti reverse protection function | ||||||
IP protection level | IP20 | |||||
Channel characteristics | Double closed loop structure is used for constant current source and constant voltage source | |||||
Channel control mode | Independent control | |||||
Voltage and current detection sampling | Four wire connection | |||||
Noise | <45dB | |||||
Data base | Apply MySQL Centralized management of test data in database | |||||
Computer communication mode | Base onTCP/IP agreement | |||||
Server operating systems | Windows 7 | |||||
Data output mode | EXCEL2003,2010、TXT | |||||
Server disk configuration | 500GB | |||||
Communication interface | Ethernet | |||||
Leakage current | 0.005uA | |||||
Machine channels | 8 | |||||
Requirements of equipment working environment | ||||||
Item | Parameter | |||||
Operating temperature range | 0℃~40℃ | |||||
(25±10℃,accuracy:0.005% of FS /℃) | ||||||
Storage Temperature Range | -10℃~50℃ | |||||
Working environment relative humidity range | ≤70% RH (No condensation of water vapor) | |||||
Storage environment relative humidity range | ≤80% RH (No condensation of water vapor) | |||||
W*D*H (mm) | 483 * 310 * 48 |