Item No.: XW-HW-200-160CH
Application: TOB New energy pouch cell and coin cell constant temperature test Chamber.
Order(MOQ): 1 set;
Payment: T/T, Western Union, Paypal, L/C;
Product Origin: China;
Lead Time: 15days;
Note: XIAOWEI factory supports OEM.

 Products Description 

XW New energy pouch cell and coin cell constant temperature test Chamber.

 Technical Parameters 

1. Product applicationConstant temperature test of polymer battery and button battery.
Manufacturers of electronics, electrical engineering, instrumentation, materials, semiconductors, etc. conduct constant temperature tests on non-flammable and non-explosive items.
Environmental protection, agricultural and livestock, aquatic scientific research institutions and production companies, etc., conduct constant temperature tests on water analysis, bacteria, mold, microbial cultivation, preservation, plant cultivation, and breeding experiments.
2. Sample limitThis test equipment prohibits:
Test or storage of flammable, explosive and volatile substances
Test or storage of corrosive material samples
Test or storage of strong electromagnetic emission source samples
Testing and storage of radioactive material samples
Test and storage of highly toxic substances
Test or storage of samples that may produce the above-mentioned substances or objects during testing or storage
3. Capacity, Dimensions, Weight
3.1 Inner capacity200L
3.2 Inner dimensionsW500*D500*H800 mm
3.3 Overall dimensionsW700*D1200*H1500 mm 
3.4 Net weightAbout 220kg
4.Performance
4.1Testing environmentThe ambient temperature is + 25℃, the relative humidity is ≤ 85%, and there is no sample in the test chamber (no load)
4.2 Temperature range0~60℃
4.3 Temperature fluctuation≤1℃ (no load and stable temperature)
4.4 Temperature deviation±2.0℃ (no load and stable temperature)
4.5 Heating time25℃→60℃ ≤30 min
4.6 Cooling time25℃→0℃ ≤50 min
5. Structure
5.1 Insulation structureOuter wall material: high-quality cold rolled steel plate, surface spray and paint treatment
Inner wall material: stainless steel plate SUS304
Box insulation material: polyurethane foam (insulation thickness 50mm)
5.2 Air conditioning channelAxial fan, heater, evaporator
5.3 Standard configurationDoor: Hollow anti-fog tempered glass + stainless steel frame
Lead hole (with soft rubber plug): φ50mm, 4pcs (located on the back side of the box)
Casters: 4 (with brake)
Sample rack: 4 layers of electrical insulation sample rack, load bearing (uniform distribution): 10kg/layer
Lighting: LED lighting
5.4 Control panelTouch control buttons
5.5 HeaterHeater control mode: contactless, equal period pulse-width modulation, SSR (solid state relay)
6.Refrigeration system
6.1 Refrigeration compressorFully enclosed piston compressor
6.2 Cooling methodAir-cooled
6.3 Throttle deviceCapillary tube
6.4 RefrigerantR134a
6.5 Welding processNitrogen-filled protective welding
7. Electrical control system
7.1 ControllerLED digital display + touch key controller
7.2 Setting modeTouch key
7.3 Control modeForced circulation ventilation. According to the set temperature value, the control system controls the output of the heater through PID automatic calculation, so as to achieve dynamic balance.
7.4 Communication modeEthernet standard interface
7.5 Temperature control moduleIndependent research and development (after high and low temperature impact, vibration and EMC and other related tests)
8.Safety protection
Test chamberLeakage protection, short circuit protection and abnormal operation of circulating fan, etc.
9.Other configuration
9.1 Power cable(single phase + protective earth wire) 1 cable
9.2 Main power leakage circuit breakerSingle phase + protective earth wire
10. Transport(the test chamber is integral, transported as a whole)
DimentionsMaximum size (without packaging): W700 mm×D1200 mm×H1500 mm 
11. Conditions of use
11.1 Installation siteThe ground is flat, with flatness ≤ 5mm / 2m
Good ventilation
There is no strong vibration around the equipment
There is no strong electromagnetic field around the equipment
There are no flammable, explosive, corrosive substances and dust around the equipment
Proper space for use and maintenance shall be reserved around the equipment, and the opening and closing space of the box door shall be reserved at the opening door of the equipment, and there shall be no other objects in front of the box door of the equipment.
11.2 Environment conditionTemperature: 5℃~ 35℃, Relative humidity: ≤85%, Air pressure: 86kpa ~ 106kpa
11.3 Power supplyAC(220±22)V, (50±0.5)Hz, single phase + protective earth wire
The grounding resistance of the earth wire is less than 4Ω
The user is required to configure the corresponding air or power switch for the equipment at the installation site, and the switch must be used independently for the equipment.
Power: 6kW
Max current: 30A
11.4 OthersOpening the door during the test will cause the temperature fluctuation in the chamber.
12.Battery specifications and placement
12.1 Battery specificationsButton battery or polymer battery
12.2 Battery placementFour layers (up to 40 button batteries can be placed on each layer)